型号 |
检测气体 |
监测范围 |
检测误差 |
重复性 |
零点漂移 (月) |
响应时间
T90 |
工作温度 (℃) |
环境湿度 (RH) |
传感器寿命 (月) |
SK6100 |
O2 * |
0∽25%VOL |
≤±0.5%VOL |
<1% |
<0.5% |
<20s |
-20∽40 |
<90% |
18 |
SK6100 |
CO* |
0∽1000ppm** |
≤±5%FS |
<2% |
<2% |
<40s |
-20∽60 |
<90% |
24 |
SK6100 |
H2S |
0∽200ppm** |
≤±5%FS |
<2% |
<2% |
<40s |
-20∽60 |
<90% |
24 |
SK6100 |
H2 |
0∽2000ppm** |
≤±5% FS |
<2% |
<2% |
<40s |
-20∽40 |
<90% |
24 |
SK6100 |
NH3 |
0∽50ppm** |
≤±10% FS |
<10% |
<5-10% |
<160s |
-20∽30 |
<90% |
24 |
SK6100 |
CL2 |
0∽20ppm** |
≤±8% FS |
<2% |
<3% |
<100s |
-20∽40 |
<90% |
24 |
SK6100 |
NO2 |
0∽20ppm** |
≤±5% FS |
<2% |
<2% |
<40s |
-20∽40 |
<90% |
24 |
SK6100 |
HCL |
0∽000ppm** |
≤±8% FS |
<2% |
<3% |
<120s |
-20∽40 |
<90% |
24 |
SK6100 |
HCN |
0∽100ppm** |
≤±8% FS |
<2% |
<3% |
<80s |
-20∽40 |
<90% |
24 |
SK6100 |
SO2 |
0∽100ppm** |
≤±5% FS |
<2% |
<2% |
<30s |
-20∽40 |
<90% |
24 |
*为可以带显示,其工作温度最低为-15℃
**为可提供其它检测范围
检测原理: |
电化学 |
◎连接电缆: |
两芯屏蔽电缆截面2x1.5mm |
检测方式: |
扩散式采样 |
◎连接距离: |
>1000m |
工作方式: |
固定式安装连续工作 |
◎安装方式: |
墙壁式( A型);管装式(B型) |
工作电压: |
12∽30V DC |
◎重量: |
1200g |
工作电流: |
50Ma(Max) |
◎关联设备: |
安全栅(推荐使用KAS9021及LB801) |
输出信号: |
4∽20Ma标准信号两线制输出 |
◎防爆形势: |
本质安全型 |
外形结构: |
压铸铝壳体,外表喷塑 |
◎防爆等级: |
IallCT6 |
外形尺寸: |
120x120x95mm |
◎防爆证号: |
CE970333 CE970334 |
![]() |
订货时在SK6100后,可选后缀A、B、D三个参数,
其中:A=墙壁是安装;B=管装式安装;C=带显示型
如:SK-6100-CO-AD 表示带显示一氧化碳探头,安装方式为墙壁式安装 |